Japan Contact Instrumentation

Export Division for Showa Measuring Instruments Co., Ltd., Japan

Home       Guide to Instrumentation       Ordering& Inquiries

Features for N11-MA-5-120-11 Strain Gauge

*Almost no effect on test object.

*Distant and multi-point measurements are possible.

*Applicable to both static and dynamic strains.

*Both surfaces being laminated, the gauge grids are entirely protected.

*The gauges, being fitted with leads, are easy to handle.

 

Standard Specifications

          Gauge Length                                   0.3mm. minimum to 60mm. maximum

          Measurable Strain                           2~4% maximum.  Up to 10% with foil-yielding strain gauges.

          Temperature Range                          FA (Polyester Backing)  -30o C to +80o C

                                                                       MA (Polyimide Backing) -30o C to +180o C

           Thermal Output(fig.1)                      FA: Within +/- 2/o C (At room temperature up to +80o C)

                                                                       MA: Within +/- 2/o C (At room temperature up to +160o C) 

                                                                                Within +/- 5/o C (At +160oC up to  +180o C)

           Gauge Factor Change with Temp.(fig.2)                                       Within +/- 0.015%/ o C  

           Gauge Resistance Tolerance            Within +/- 0.5% of the nominal resistance.

           Gage Factor                                         2.00 (nominal) 

          Gage Factor Tolerance                        Within +/- 1% of the value indicated on individual gauge

                                                                                        packet for gauge lengths of 5mm. to 60mm.

                                                                          Within +/- 2% of the value indicated on individual gauge

                                                                                         packet for gauge lengths of 0.3mm. to 3mm.

          Fatigue life                                             More than 105 reversals at 1000x10-6 strain.

            

Apparent Strain  (x 10-6)

 

 

 

 

 

 

 

Gauge Factor

General Performances of Type N11-MA-5-120-11 Strain Gauge

Thermal Output Characteristics (Fig. 1)

Gauge Factor Variation with temperature (Fig. 2)

Configurations

N11 

Basic Pattern and  its Combinations 

 

F

Base Material

F: Polyester       M:  Polyimide

 

A

Foil Material

A:  Cu-Ni Alloy

 

5

Gauge Length

Expressing grid effective length in figures in the unit of mm.

 

350

Gauge Resistance

Expressing strain gauge          nominal resistance in the unit of Ω.                                                             

Can be deleted when nominal resistance is 120Ω.

 

16

Linear Expansion Factor of material against which strain                                   gauge is self-temperature compensated and its base color  compensated.    (see chart 1 below)

 

L03

Optional Specifications  

03  Length of Leads

L    Resin Clad Copper Wire (Up to 0.3m)

W   Waterproof Molded Type

P    Optional Pattern Types (subject to special quotations)

 

Load Cell & Strain Gauge Index

TOP

Orders & Inquiries