Troubleshooting Guide For Load Cells |
| UNAMPLIFIED TRANSDUCERS | |
| SYMPTOM/PROBLEM |
ACTION |
| No Output |
Verify correct wiring. |
| Verify excitation per calibration sheet. | |
| Make sure pressure, load, etc. is being applied. | |
| Check strain gage bridge for continuity. | |
| Erratic/Intermittent |
Check electrical connection for discontinuity or damage. |
| Output or Zero Drift |
Check for isolation resistance between bridge wiring and transducer body. |
| Make sure pressure, load, etc. is constant. | |
| Check stability of excitation power supply. | |
| Check millivolt output with voltmeter. | |
| Check for RFI/EMI interference. | |
| Make sure there are no rapid changes in temperature. | |
| Incorrect Output |
Check actual input and output resistance for possible change or open bridge. |
| Check zero offset to see if high. | |
| High Zero Offset |
Usually indicates transducer was over ranged beyond specifications. Over ranged transducers should be recalibrated. |
| Check actual input and output resistance measurements for possible changes or open bridge. |
|
| TRANSDUCER WITH INSTRUMENT READOUT | |
| SYMPTOM/PROBLEM |
ACTION |
| Erratic Display |
Check electrical connection and wiring code for pin layout. |
| No Display/No Output Voltage |
Check power line fuse per instrument instructions. |
| Blinking Display |
Indicated overload; make sure wires are all connected and transducer is within its range. |
| Incorrect Readout |
Check transducer range on label. |
| Verify that system was set-up per instructions. Review set-up procedure. |
|
| Verify that transducer operates properly. | |
| Use Resistor Shunt-Calibration to verify calibration when integrated with Instrument Readout, or connect resistor across (-) Excitation and (-) Signal. Refer to Certificate of Calibration for correct resistor value. |
|