Troubleshooting Guide For Load Cells

 
UNAMPLIFIED TRANSDUCERS
SYMPTOM/PROBLEM ACTION
No Output Verify correct wiring.
  Verify excitation per calibration sheet.
  Make sure pressure, load, etc. is being applied.
  Check strain gage bridge for continuity.
Erratic/Intermittent Check electrical connection for discontinuity or damage.
Output or Zero Drift Check for isolation resistance between bridge wiring
and transducer body.
  Make sure pressure, load, etc. is constant.
  Check stability of excitation power supply.
  Check millivolt output with voltmeter.
  Check for RFI/EMI interference.
  Make sure there are no rapid changes in temperature.
Incorrect Output Check actual input and output resistance for possible change or open bridge.
  Check zero offset to see if high.
High Zero Offset Usually indicates transducer was over ranged beyond
specifications. Over ranged transducers should be
recalibrated.
  Check actual input and output resistance measurements
 for possible changes or open bridge.
  TRANSDUCER WITH INSTRUMENT READOUT
SYMPTOM/PROBLEM ACTION
Erratic Display Check electrical connection and wiring code for pin layout.
No Display/No Output Voltage Check power line fuse per instrument instructions.
Blinking Display Indicated overload; make sure wires are all connected and
transducer is within its range.
Incorrect Readout Check transducer range on label.
  Verify that system was set-up per instructions. Review
set-up procedure.
  Verify that transducer operates properly.
  Use Resistor Shunt-Calibration to verify calibration when
integrated with Instrument Readout, or connect resistor
across (-) Excitation and (-) Signal. Refer to Certificate
of Calibration for correct resistor value.